Home > Products > Elemental Analyzer

5E Series C/H/N Elemental Analyzer

  • CKIC 5E Series C/H/N Elemental Analyzer
  • CKIC 5E Series C/H/N Elemental Analyzer
  • CKIC 5E Series C/H/N Elemental Analyzer
  • CKIC 5E Series C/H/N Elemental Analyzer

    5E Series C/H/N Elemental Analyzer is used to determine carbon, hydrogen, nitrogen content in solid and liquid material, such as coal, coke, oil, petroleum, biomass, fertilizer, plastic, food, hydrocarbons and plant tissue, leaves and tobacco, which is widely applied in power plants, coal mines, metallurgy, chemical industry, commercial inspection, scientific research, food industry, education etc.

    Please leave your E-mail address before download the product PDF. Your cooperation is appreciated.

    Warm prompt: Please make sure you entered a correct E-mail address. We will send you relevant information and product quotation regularly.

    • Features
    • Specifications
    • Spare parts
    • Case
    Standard Configuration: Computer
    Main analyzer
    Furnace reagent
    High purity copper
    N-Catalyst
    H2O sorb reagent
    Standrad Reference Material(GBW)

    Optional Configuration:
    AR427 com-aid for liquid sample
    Additional1 2 4 layers carousels
    4cm x 4cm size tin-foil cup
    Bigger size hole carousel

    Models Available:
    5E-CHN2200 to test Carbon, Hydrogen, Nitrogen content
    5E-CH2200 to test Carbon, Hydrogen content
    5E-TCN2200 to test Nitrogen content
    5E-TCN2210 to test Nitrogen contentPre-sale
    5E-IRH2200 to test Hydrogen content
    5E-IRC2200 to test Carbon content
    5E-CN2200 to test Carbon content, Nitrogen content

    CO2 sorb reagent

    Silica wool 
    Lower crucible
    Upper crucible
    O-ring kit
    Tool kit







    Features:
    Maximum Efficiency
    1. High throughput: standard auto loader for 35 samples per layer, stackable to 4 layers available.
    2. Dual-stage furnace system with pure oxygen flow to ensure the complete combustion of all samples.

    Good Environment Adaptability 
    1. Optimum gas circuit provides good gas tightness of the system.
    2. O-ring free from heat resource.

    Minimum Consumption
    1. Independent detectors to determine C, H, N respectively. Analysis of CH mode and CHN mode can be chosen on software. (For 5E-CHN2200)
    2. Saving time, gas and reagent: only 5.5ml blended gas needed to be analyzed.

    Unattended Operation
    The operator is limited to just adding sample to auto sample loader. Then the instrument will finish the test, cool down and shut off automatically.

    Model 5E-CHN2200 Elemental Analyzer
    Conforms to Method AS 1038.6.4, ASTM D5373, ASTM D5291, ISO 16634, ISO 16948, ISO 29541, EN 15407,EN ISO 16948:2015-0, GB/T 30728, GB/T 30733, UNI 15104, UNI CEN/TS 15407
    Analysis Time 7 mins, depending on sample combustion conditions
    Sample Loader Stackable auto loader, up to 140 samples by 4 layers
    Repeatability Carbon(Cad)≤0.5%, Hydrogen(Had)≤0.15%, Nitrogen(Nad)≤0.08%
    Sample Mass Up to 1000mg, depending on sample matrix
    Temp. Resolution 1
    Gas Required* Helium, 99.99%, 0.25± 0.01 Mpa
    Oxygen, 99.99%, 0.25± 0.01 Mpa
    Nitrogen or compressed air, 0.25± 0.01 Mpa
    Consumption Helium 200ml/min
    Furnace Type Resistance furnace, Max temp 1050
    Power Supply Single phase, AC220V ±10% , 50/60Hz, 5.5kW
    Net Weight 110kg
    Dimensions(L×W×H) 690mm×750mm×720mm
    *Test Condition: If N≤0.8%, 99.999% high purity oxygen is required.

    InquiryPlease feel free to submit your inquiry information to us.